TM 11-6625-3041-30/TO 33A1-8-908-12
2-6.
SG-1139 CHECK PROCEDURE (CONT)
Step
This step checks the following:
16
4
5
Receive 2 card . . . BER Duration Counter 10 and 2 x 10
cont
outputs
Red Error Count 10 output
Yellow Error Count 2 output
If any BER indication is improper:
Remove LOGIC access cover (refer to TM 11-6625-3041-12) and
note which card contains a lit fault indicator. Replace that
card and repeat step 16.
If no card contains a lit fault indicator, possible cause is
Receive 2 card A6. Replace (refer to TM 11-6625-3041-12), and
repeat step 16.
END
2-26