2-16. POWER CABLE REPAIR (CONT)
Male
Connector
Replacement
the same length.
Strip cable insulation so that it is 1 inch back from end
of wires.
Strip wire insulation so that it is 0.5 inch back from end
of wires.
Twist end of each wire.
For assembly, see Male Connector Assembly below.
Use flat-tip screwdriver to loosen all three contact screws.
Pull wires out of contacts.
Replace damaged or missing connector Components.
Male
Connector
Assembly
Insert wires in contacts.
Use flat-tip screwdriver to tighten all three contact screws.
2-38