TM 11-6625-3041-12/TO 33A1-8-908-1
SELF-TEST (cont)
2-4.
Press DISPLAY
Step .6-32
q
RESET, then press
6
No-Go
SINGLE ERROR
Detection
rapidly 10 times
(all presses must
be done within 15
seconds).
ERRORS display must
read 10 and red NO GO indicator must light.
After 5 to 6 minutes, NO GO indicator will extinguish and GO
indicator will light. You need not wait for this, however,
and may proceed immediately to the next step.
Set DATA RATE to each of the remaining positions (1.2 thru
Step .6-32
q
32):
Remaining
7
Rates
At each position,
DATA IN and TMG
IN indicators must
remain lit. FAULT
indicator must not
be lit.
Keep DATA RATE at
Step .6-32
q
32, press RESET
Marginal
8
DISPLAY and press
and GO
SINGLE ERROR two
Detection
times (all presses
must be done within
5 seconds).
ERRORS display must
read 2 and yellow MARGINAL indicator must light.
After 6 to 7 seconds,
MARGINAL indicator
must extinguish and
GO indicator must
then light.
Disconnect cable and adapters from DIPHASE IN and DIPHASE OUT.
q
2-14