TM 11-6625-3041-12/TO 33A1-8-908-1
3-11. TROUBLESHOOTING PROCEDURE (cont)
If STA CLOCK indicator is lit before LAMP TEST is pressed:
Possible cause is Multi VDC card Al (-12V supply defective).
IF STA CLOCK indicator does not light:
Possible cause is Transmit 1 card A3.
Replace (refer to
If replacement Multi VDC card Al does not provide proper indi-
cation, possible cause is STA CLOCK indicator lamp. Evacuate
to direct support for repair.