TM11-6625-2948-14+P
NOTE
MODULATION OPERATION
Undelayed or delayed sync out signals are
available at the DELAYED SYNC OUT con-
as follows:
nector and/or the UNDELAYED SYNC
OUT connector of the modulator.
a. Set the FUNCTION switch of the modulator to EXT
lNT
(Internal to SG-1145/URM)
b. Set the SELECTOR switch of the modulator to INT.
c. Set the MODULATION SELECTOR switch of the signal
as follows:
generator to EXT
a. Set the MODULATION SELECTOR switch of the signal
d . Adjust the RATE control of the modulator for the
generator to INT FM ~.
desired rate.
b. Adjust the INT FM DEV control to obtain the desired
sweep width.
INTERNAL PULSE MODULATION OPERATION
as follows:
as follows:
a. Set the FUNCTION switch of the modulator to lNT
a. Set the MODULATION SELECTOR switch of signal
b. Set the SELECTOR switch of the modulator to INT.
c. Set the MODULATION SELECTOR switch of the signal
b. Set the FUNCTION switch to FM and the SELECTOR
generator to EXT
switch to INT on modulator MD-1075/URM.
d. Adjust the RATE and PULSE WIDTH controls of the
c . Adjust the RATE control and multiplier switch X1,
modulator for the desired pulse rate and width.
X10, X100 on modulator MD-1075/URM for the desired
modulation rate.
d . Adjust the FM DEV AMPLITUDE control for the
desired deviation.
as follows:
a. Set the FUNCTION switch of the modulator to EXT
/
INT
a. Connect the external modulating signal to the EXT FM
connector.
b. Select the SELECTOR switch of the modulator to
EXT(+) or EXT(-) as determined by the externally applied
pulse polarity.
c. Set MODULATION SELECTOR switch of the signal
c. Set the MODULATION SELECTOR switch of the signal
--
generator to EXT
d. Set FUNCTION switch of the modulator to OFF.
d. Connect the external modulating signal to the EXT
INPUT connector of the modulator.
e . Set the external FM source to desired characteristics.