TM 11-6625-3053-14
Make set-up as shown below.
Energize signal generator and wait 2 minutes.
Set MODULATION switch to EXTERNAL PULSE.
Set to desired frequency and set output level or ALC operation.
Energize pulse generator.
Set pulse generator to desired frequency and amplitude.
i.
External AM Modulation:
Auxiliary equipment required: 1 oscillator capable of producing a sine wave signal:
Frequency: 0 to 1 MHz.
Amplitude: 5 to 6V peak to peak.
HP Model 652A or equivalent.
Make set-up as shown below.
Set signal generator to initial adjustments of para 2-4.
Energize signal generator and wait 2 minutes.
Set MODULATION switch to EXTERNAL AM.
Set to desired frequency and set output level, or ALC.
Set oscillator to desired frequency and amplitude.
2-15