TM 11-6625-3053-14
ITEM
MALFUNCTION
CHECKS/CORRECTIVE ACTION
6
b) Remove high voltage cover if on (para 6-22).
(cont'd)
c) Set multimeter to 1000 VDC range.
d)
e) Energize sig gen and multimeter.
f)
Wait 2 mins and set MODULATION Switch to CW.
g) Using multimeter, check voltage between chassis ground and test
pin 12 on A2A1 card assy. Voltage should read -325 VDC.
If voltage is not obtained, go to T/S item 7.
If voltage is obtained, go to check 2 below.
2. Check -425V output from A2A1 card assy.
a) Using multimeter as set, measure voltage between test pin 12
and test pin 5 on the A2A1 card assy.
Voltage should be -425VDC.
If voltage obtained, check wiring between test pin 5 on A2A1 card
and -750V test point on tracking board assy. Repair as required.
If voltage not obtained, go to check 3 below.
a) Using multimeter, set to 1000 VDC range and measure voltage
between test pin 12 and junction of C1 and R12 on A2A1 card
assy.
Voltage should be greater than -495 VDC.
If voltage is not obtained, go to check 4 below.
If voltage is obtained, check R12 and CR5. Replace faulty
component.
If this does not correct the -425 output, go to check 5 below.
4. Input check.
a) Using multimeter, set to 1000 VAC range. Measure voltage
across test pins 1 and 2 on A2A1 card assy.
Voltage should be approx 593 VAC.
If voltage not obtained, go to troubleshooting item 8.
If voltage obtained, check CR1 through CR4 and replace
defective component.
6-56